Cardinal Health 200, LLC: Medical Device Recall in 2017 - (Recall #: Z-2146-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Jackson-Pratt Silicone Round Drain, 10FR. Sterile, Rx Only. For Single Use Only. Jackson-Pratt Wound Drainage Systems are sterile single use devices. The wound drainage system consist of a wound drain and a fluid collection reservoir. The wound drain is a radio-opaque section of tubing which has perforations or ducts. The wound drain is surgically placed in a surgical wound site using a trocar. Blood and fluids are collected through a wound drain into a fluid collection reservoir. Wound drains are used to remove exudates from wound sites. Trocars, drains and reservoirs are for a single patient use only and should be discarded after use
Class II
Product's seal possibly compromised, potentially compromising the sterility of the package contents. Use of impacted products could result in an increased risk of infection.
Catalog Number: SU130-1321 Lot Numbers: 1161206 1161207 1161370 1161371 1161372 1161476 1161477 1161478 1161606 1161607 1161608 1170036
Nationwide Distribution
Voluntary: Firm initiated