Cardinal Health 200, LLC: Medical Device Recall in 2020 - (Recall #: Z-0424-2021)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Covidien MONOJECT Standard Hypodermic Needle, Polypropylene Hub 22 G x 3/4". Intended use to facilitate activities such as: withdrawal of medication from a container, aspiration of fluid from a patient, injection of medication into a patient, and transfer fluids Item Code: 8881250248
Class II
Potential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not securely seated, the needle can fall out of the sheath and cause the unused needle to project out of the cartridge, and result in a needle stick/puncture from the clean/unused needle
Lot Number: 011144 Expanded Recall 11/4/20: Lot Number: 011145
Distribution US Nationwide, Australia and Canada
Voluntary: Firm initiated