Cardinal Health 200, LLC: Medical Device Recall in 2020 - (Recall #: Z-1453-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Monoject 3mL Syringe with Hypodermic Safety Needle, 22G x 1-1/2 . Item Code 11832215 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.
Class II
Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.
Lot No. 15072015, Item Code 11832215 , Expires 06/20/2020.
Worldwide distributions - US Nationwide distributions in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and countries of Canada, Australia, South Africa, Belgium, Germany, Ireland, Portugal Spain, and UK.
Voluntary: Firm initiated