Cardinal Health Inc.: Medical Device Recall in 2021 - (Recall #: Z-1933-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
BD UF PEN 32GX4MM 100 NANO Item Number: 4292272
Product Classification:
Class II
Date Initiated: March 16, 2021
Date Posted: June 30, 2021
Recall Number: Z-1933-2021
Event ID: 87787
Reason for Recall:
Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
Status: Ongoing
Product Quantity: 1,599 boxes of 100 needles
Code Information:
NDC/UPC: 8290320122 All lots numbers shipped from 2/23/201 to 3/10/2021
Distribution Pattern:
US Nationwide distribution in the states of FL, GA, SC.
Voluntary or Mandated:
Voluntary: Firm initiated