Cardinal Health: Medical Device Recall in 2013 - (Recall #: Z-0978-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 3FA40SSA) Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 3FM40SSA) Product Usage: Disposal system for counting of used surgical blades and needles.
Class II
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
Catalog Number: 3FA40SSA Lot Numbers: all lot numbers less than or equal to 928782 Catalog Number: 3FM40SSA Lot Numbers: all lot numbers less than or equal to 911379
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Voluntary: Firm initiated