Cardinal Health: Medical Device Recall in 2013 - (Recall #: Z-0980-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FS30SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Double Foam Strips (Cat. No. 4FS31SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Double Numbered Magnets (Cat. No. 4MM30SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

Product Classification:

Class II

Date Initiated: December 18, 2012
Date Posted: April 3, 2013
Recall Number: Z-0980-2013
Event ID: 63951
Reason for Recall:

Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

Status: Terminated
Product Quantity: Cat. No. 4FS30SSA: 7488; Cat. No. 4FS31SSA: 48; Cat. No. 4MM30SSA: 6336
Code Information:

Catalog Number: 4FS30SSA Lot Numbers: all lot numbers less than or equal to 848031 Catalog Number: 4FS31SSA Lot Numbers: lot number 739269 Catalog Number: 4MM30SSA Lot Numbers: all lot numbers less than or equal to 858154

Distribution Pattern:

Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.

Voluntary or Mandated:

Voluntary: Firm initiated