Cardinal Health: Medical Device Recall in 2013 - (Recall #: Z-0980-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FS30SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Double Foam Strips (Cat. No. 4FS31SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Double Numbered Magnets (Cat. No. 4MM30SSA) Product Usage: Disposal system for counting of used surgical blades and needles.
Class II
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
Catalog Number: 4FS30SSA Lot Numbers: all lot numbers less than or equal to 848031 Catalog Number: 4FS31SSA Lot Numbers: lot number 739269 Catalog Number: 4MM30SSA Lot Numbers: all lot numbers less than or equal to 858154
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Voluntary: Firm initiated