Carefusion 2200 Inc: Medical Device Recall in 2020 - (Recall #: Z-2400-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.
Class II
Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.
Model Numbers: OP0702 and OP0703. Model OP0702, Lot C20VAA ( 4 inch length) Model OP0703, Lot D17VAA (6 inch length)
US Nationwide distribution including in the states of CA, FL, VA and KY.
Voluntary: Firm initiated