Carefusion 2200 Inc: Medical Device Recall in 2020 - (Recall #: Z-2655-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Safe-T 8 Fr Thora-Para Tray With 8 Fr. Catheter and Sharps Injury Protection - Drug Free
Class II
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Catalog: TPT1000SDF UDI: 50885403108493; Batches: 0001320515, 0001321729, 0001323273, 0001323703, 0001332819, 0001334405, 0001338453, 0001344142, 0001344740
370 US customers and 13 distributors. Product distributed in: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MD, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY Customer listed, including Federal distribution, uploaded.
Voluntary: Firm initiated