CareFusion 303, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2741-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Alaris System PC Unit Model 8015. modular infusion pump and monitoring system

Product Classification:

Class I

Date Initiated: June 30, 2020
Date Posted: August 12, 2020
Recall Number: Z-2741-2020
Event ID: 85915
Reason for Recall:

If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.

Status: Terminated
Product Quantity: 824277
Code Information:

All serial numbers

Distribution Pattern:

U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, BH

Voluntary or Mandated:

Voluntary: Firm initiated