CareFusion 303, Inc.: Medical Device Recall in 2025 - (Recall #: Z-1375-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC, REF: 1156-00, CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01, CII Safe ES Desktop PC, W-less SCNR, REF: 107-255-01, Used with Enterprise Server (ES), REF: 1115-00 with affected Enterprise Server software.
Class II
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
BD Pyxis Enterprise Server software versions: 1.7.4.128, 1.7.4.134, 1.7.4.136, 1.7.4.137, All 1.7.1, All 1.7.2, All 1.7.3. REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674, 1116-00/10885403512605
Worldwide - US Nationwide distribution including in the states of KY, OH, NY, MO, NC, PA, FL, TN, CT, LA, MN, VA, MD, CA, TX, NE, IA, OK, MI, PR, SD, IN, ME, KS, IL, NM, WA, AR, HI, MS, GA, WY, WI, VT, AL, NV, NJ, DC, AZ, MT, AK, OR, CO, NH, MA, SC, ID, UT, ND and the countries of TW, CA, AE, MX, SG, AU, NZ, GB, CH, BH.
Voluntary: Firm initiated