CareFusion 303, Inc.: Medical Device Recall in 2025 - (Recall #: Z-1639-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

BD PYXIS MEDBANK (Name, REF): MEDPASS MODULE, 139088-01; MINI 1FH-1FM, 169-137; MINI 1HH-1FH-1HM, 169-101; MINI 1HH-1HM-1FM, 169-102; MINI 2FH, 138902-01; MINI 2FH, 169-104; MINI 2FM, 169-105; MINI 2HH-1FH, 169-106; MINI 2HH-1FH, 138908-01; MINI 2HH-1FM, 169-107; MINI 2HH-1FM, 138906-01; MINI 3HH-1FG, 169-110; MINI 3HH-1FG, 139091-01; MINI 3HH-1HM, 169-111; MINI 4HH, 138910-01; MINI 4HH, 169-112; MINI 4HM, 169-113; MINI 4HM, 138915-01; TWR MN 10HH-1FH-1FM, 169-23; TWR MN 10HH-2FH, 169-24; TWR MN 10HH-2FH, 138955-01; TWR MN 10HH-2FM, 169-25; TWR MN 10HH-2FM, 138924-01; TWR MN 12HH-1FF, 169-26; TWR MN 12HH-1FF, 138960-01; TWR MN 12HH-1FH, 169-27; TWR MN 12HH-1FH, 138961-01; TWR MN 12HH-1FM, 169-28; TWR MN 12HH-1FM, 138926-01; TWR MN 14HH, 138927-01; TWR MN 14HH, 169-29; TWR MN 1HH-1HM-2FM-4FF, 169-30; TWR MN 2FH-2HM-4FM, 169-31; TWR MN 2HH-10HM-1FM, 169-32; TWR MN 2HH-1FH-5FM, 169-33; TWR MN 2HH-2FH-4FM, 169-34; TWR MN 2HH-4FH-2FM, 169-139; TWR MN 2HH-4HM-4FM, 139125-01; TWR MN 2HH-6FM, 169-38; TWR MN 2HH-6FM, 138919-01; TWR MN 2HH-6HM-3FM, 169-39; TWR MN 2HH-8HM-2FM, 169-40; TWR MN 2HM-6FM, 169-41; TWR MN 3FH-4FM, 169-42; TWR MN 3FH-4FM, 138933-01; TWR MN 3FH-8HH MEDPASS, 169-140; TWR MN 3HH-7HM-2FM, 169-43; TWR MN 4HH-10HM, 169-44; TWR MN 4HH-10HM, 138959-01; TWR MN 4HH-1FH-2HM-3FM, 169-45; TWR MN 4HH-1FH-4FM, 169-46; TWR MN 4HH-1FH-4HM-2FM, 169-47; TWR MN 4HH-2FH-2HM-2FM, 169-49; TWR MN 4HH-2FH-3FM, 169-50; TWR MN 4HH-2FH-3FM, 138931-01; TWR MN 4HH-3FH-2FM, 169-52; TWR MN 4HH-4FH-1FM, 169-53; TWR MN 4HH-4FH-1FM, 138940-01; TWR MN 4HH-4HM-3FM, 169-54; TWR MN 4HH-4HM-3FM, 138952-01; TWR MN 4HH-5FH, 169-55; TWR MN 4HH-5FM, 169-56; TWR MN 4HH-5FM, 138918-01; TWR MN 4HH-6HM-2FM, 169-57; TWR MN 4HM-5FM, 169-59; TWR MN 5HH-3HM-3FM, 169-60; TWR MN 5HH-9HM, 138914-01; TWR MN 5HH-9HM, 169-61; TWR MN 6FH-1FM, 169-62; TWR MN 6HH-1FH-2HM-2FM, 169-63; TWR MN 6HH-1FH-3FM, 169-64; TWR MN 6HH-2FH-2FM, 169-65; TWR MN 6HH-4FH, 169-67; TWR MN 6HH-4FH, 138957-01; TWR MN 6HH-4FM, 169-68; TWR MN 6HH-4FM, 138922-01; TWR MN 6HH-4HM-2FM, 169-69; TWR MN 6HH-8HM, 169-71; TWR MN 6HH-8HM, 138958-01; TWR MN 7FH, 169-72; TWR MN 7FH, 138920-01; TWR MN 7FM, 169-73; TWR MN 7FM, 138937-01; TWR MN 7HH-1HM-3FM, 139086-01; TWR MN 7HH-1HM-3FM, 169-74; TWR MN 8HH-1FH-2FM, 169-75; TWR MN 8HH-2FH-1FM, 169-77; TWR MN 8HH-2HM-2FM, 169-78; TWR MN 8HH-2HM-2FM, 138939-01; TWR MN 8HH-3FF, 169-79; TWR MN 8HH-3FH, 169-80; TWR MN 8HH-3FH, 138956-01; TWR MN 8HH-3FM, 169-81; TWR MN 8HH-3FM, 138923-01. BD PYXIS MEDFLEX MN (Name, REF): 1000 12HH, 139055-01; 1000 2HH-2FM, 139041-01; 1000 2HH-5FM, 139046-01; 1000 4HH-4FM, 139049-01; 1000 6HH-3FM, 139051-01; 1000 8HH-2FM, 139052-01; 2000 8HH-2FH, 139165-01; 2000 8HH-2FM, 139069-01. MEDBANK MAIN, 1137-00; MEDBANK MINI MAIN, 1147-00; MEDBANK TOWER MAIN, 1145-00; MEDFLEX, 1119-00; MEDFLEX 2.0, 1139-00

Product Classification:

Class II

Date Initiated: March 11, 2025
Date Posted: April 30, 2025
Recall Number: Z-1639-2025
Event ID: 96564
Reason for Recall:

Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.

Status: Ongoing
Product Quantity: 7,357
Code Information:

REF/UDI-DI(GTIN)/Software Version: 139088-01/1.2.3.23; 169-137/10885403512544/2.4; 169-101/10885403512544/2.4; 169-102/10885403512544/2.4; 138902-01/2.4; 169-104/10885403512544/2.4; 169-105/10885403512544/2.4; 169-106/10885403512544/2.4; 138908-01/2.4; 169-107/10885403512544/2.4; 138906-01/2.4; 169-108/10885403512544/2.4; 169-110/10885403512544/2.4; 139091-01/2.4; 169-111/10885403512544/2.4; 138910-01/2.4; 169-112/10885403512544/2.4; 169-113/10885403512544/2.4; 138915-01/2.4; 169-23/10885403512568/2.4; 169-24/10885403512568/2.4; 138955-01/2.4; 169-25/10885403512568/2.4; 138924-01/2.4; 169-26/10885403512568/2.4; 138960-01/2.4; 169-27/10885403512568/2.4; 138961-01/2.4; 169-28/10885403512568/2.4; 138926-01/2.4; 138927-01/2.4; 169-29/10885403512568/2.4; 169-30/10885403512568/2.4; 169-31/10885403512568/2.4; 169-32/10885403512568/2.4; 169-33/10885403512568/2.4; 169-34/10885403512568/2.4; 169-139/10885403512568/2.4; 139125-01/2.4; 169-38/10885403512568/2.4; 138919-01/2.4; 169-39/10885403512568/2.4; 169-40/10885403512568/2.4; 169-41/10885403512568/2.4; 169-42/10885403512568/2.4; 138933-01/2.4; 169-140/10885403512568/1.2.3.23; 169-43/10885403512568/2.4; 169-44/10885403512568/2.4; 138959-01/2.4; 169-45/10885403512568/2.4; 169-46/10885403512568/2.4; 169-47/10885403512568/2.4; 169-49/10885403512568/2.4; 169-50/10885403512568/2.4; 138931-01/2.4; 169-52/10885403512568/2.4; 169-53/10885403512568/2.4; 138940-01/2.4; 169-54/10885403512568/2.4; 138952-01/2.4; 169-55/10885403512568/2.4; 169-56/10885403512568/2.4; 138918-01/2.4; 169-57/10885403512568/2.4; 169-59/10885403512568/2.4; 169-60/10885403512568/2.4; 138914-01/2.4; 169-61/10885403512568/2.4; 169-62/10885403512568/2.4; 169-63/10885403512568/2.4; 169-64/10885403512568/2.4; 169-65/10885403512568/2.4; 169-67/10885403512568/2.4; 138957-01/2.4; 169-68/10885403512568/2.4; 138922-01/2.4; 169-69/10885403512568/2.4; 169-71/10885403512568/2.4; 138958-01/2.4; 169-72/10885403512568/2.4; 138920-01/2.4; 169-73/10885403512568/2.4; 138937-01/2.4; 139086-01/2.4; 169-74/10885403512568/2.4; 169-75/10885403512568/2.4; 169-77/10885403512568/2.4; 169-78/10885403512568/2.4; 138939-01/2.4; 169-79/10885403512568/2.4; 169-80/10885403512568/2.4; 138956-01/2.4; 169-81/10885403512568/2.4; 138923-01/2.4; 139055-01/2.4; 139041-01/2.4; 139046-01/2.4; 139049-01/2.4; 139051-01/2.4; 139052-01/2.4; 139165-01/2.4; 139069-01/2.4; 1137-00/2.4; 1147-00/10885403517181/2.4; 1145-00/10885403517167/2.4; 1119-00/10885403512704/2.4; 1139-00/10885403512704/2.4

Distribution Pattern:

US Nationwide distribution including in the states of MO, MI, FL, PA, SC, OR, AZ, CO, ND, GA, NE, TX, IL, OH, VT, IN, TN, KY, IA, MT, NY, CA, UT, OK, SD, WA, AK, ME, NV, NM, NC, WV, LA, MN, ID, AL, VA, WI, NH, KS, MA, CT, RI, NJ, MD, MS, WY, DC, DE, HI, AR, ON.

Voluntary or Mandated:

Voluntary: Firm initiated