CAREFUSION: Medical Device Recall in 2021 - (Recall #: Z-2457-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

BD Syringe Set 20 DP MIC TBG 0.2MF CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798696; BD Syringe Set 20 DP MIC TBG CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798703;

Product Classification:

Class II

Date Initiated: August 2, 2021
Date Posted: September 22, 2021
Recall Number: Z-2457-2021
Event ID: 88203
Reason for Recall:

Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.

Status: Ongoing
Product Quantity: 16,700
Code Information:

Lots: 20026497 20025590 20076421 21015406 21015407

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR.

Voluntary or Mandated:

Voluntary: Firm initiated