CAREstream Medical LLC: Medical Device Recall in 2018 - (Recall #: Z-2887-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

CAREstream ProNox Nitrous Oxide Delivery System labeled as the following: a. CMI-0100 PNX, Pro-Nox Mixer - Canadian; b. CMI-0100-PNX-HS, Pro-Nox Complete Hospital Roll Stand System - Nitrous Oxide Delivery System; c. CMI-0100-PNX-HS-NO2H, Pro-Nox Hospital System without Oxygen Regulator; d. CMI-0100-PNX-HSWM, Pro-Nox Hospital System Wall Mount with Universal Mount (GCX) scavenger; e. CMI-0100-PNX-NHS, Pro-Nox Complete Roll Stand Nitrous Oxide Delivery System (Non-hospital); f. CMI-0100-PNX-NHS-PS, Pro-Nox Complete Roll Stand Nitrous Oxide Delivery System with Passive Scavenger; g. CMI-0100-PNX-US, Pro-Nox mixer, 6' high-pressure O2 and N2O hoses

Product Classification:

Class II

Date Initiated: May 15, 2018
Date Posted: August 29, 2018
Recall Number: Z-2887-2018
Event ID: 80340
Reason for Recall:

Two analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than specification during routine testing.

Status: Terminated
Product Quantity: 919 Units Total
Code Information:

Item Number CMI-0100-PNX-xxx

Distribution Pattern:

Domestic Distribution: AK, AL, AR ,AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV. International Distribution: Canada and Morocco.

Voluntary or Mandated:

Voluntary: Firm initiated