Carl Zeiss Meditec AG: Medical Device Recall in 2020 - (Recall #: Z-0613-2021)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.
Class II
Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.
Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.
Voluntary: Firm initiated