Carl Zeiss Meditec AG: Medical Device Recall in 2020 - (Recall #: Z-0614-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses.

Product Classification:

Class II

Date Initiated: August 6, 2020
Date Posted: December 23, 2020
Recall Number: Z-0614-2021
Event ID: 86366
Reason for Recall:

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Status: Terminated
Product Quantity: 132 units
Code Information:

Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending

Distribution Pattern:

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated