Carl Zeiss Metrology Inc: Medical Device Recall in 2018 - (Recall #: Z-0379-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.
Product Classification:
Class II
Date Initiated: November 14, 2017
Date Posted: February 7, 2018
Recall Number: Z-0379-2018
Event ID: 78885
Reason for Recall:
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
Status: Ongoing
Product Quantity: 13 US
Code Information:
Serial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342
Distribution Pattern:
Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.
Voluntary or Mandated:
FDA Mandated