Cenorin, LLC: Medical Device Recall in 2014 - (Recall #: Z-1995-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
Class II
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.
Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
Voluntary: Firm initiated