CHANGE HEALTHCARE CANADA COMPANY: Medical Device Recall in 2024 - (Recall #: Z-0126-2025)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
Class II
A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
UDI Numbers: (01)17540262100013(10)140100, (01)17540262100020(10)140101, (01)17540262100037(10)140200, (01)17540262100044(10)140300, (01)17540262100051(10)140302, (01)17540262100099(10)150000, (01)17540262100105(10)150001, (01)17540262100129(10)150100. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom.
Voluntary: Firm initiated