CHANGE HEALTHCARE CANADA COMPANY: Medical Device Recall in 2024 - (Recall #: Z-0127-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Horizon Cardiology Hemo. Version 12.2.
Product Classification:
Class II
Date Initiated: August 21, 2024
Date Posted: October 23, 2024
Recall Number: Z-0127-2025
Event ID: 95284
Reason for Recall:
A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
Status: Ongoing
Product Quantity: 3 systems
Code Information:
No UDI. Version 12.2
Distribution Pattern:
Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated