Childrens Medical Ventures: Medical Device Recall in 2014 - (Recall #: Z-1789-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Product Classification:

Class II

Date Initiated: April 29, 2014
Date Posted: June 25, 2014
Recall Number: Z-1789-2014
Event ID: 68226
Reason for Recall:

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

Status: Terminated
Product Quantity: 112
Code Information:

989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384

Distribution Pattern:

Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.

Voluntary or Mandated:

Voluntary: Firm initiated