Church & Dwight Inc: Medical Device Recall in 2023 - (Recall #: Z-1350-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)
Class II
Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.
All lots with the Easy Read App
US Nationwide distribution.
Voluntary: Firm initiated