Cincinnati Sub-Zero Products LLC, a Gentherm Company: Medical Device Recall in 2019 - (Recall #: Z-0825-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Hemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE Product Usage: The HEMOTHERM¿ Model 400CE Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator / Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper / hypothermia blanket under the patient to provide warming through conductive heat transfer.

Product Classification:

Class II

Date Initiated: December 19, 2018
Date Posted: February 20, 2019
Recall Number: Z-0825-2019
Event ID: 81961
Reason for Recall:

Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration on the board due to overheating and shut down of the device.

Status: Terminated
Product Quantity: 396
Code Information:

Serial numbers 151-10283CE through 171-10721CE

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, and WI. The products were distributed to the following foreign countries: Columbia and Taiwan.

Voluntary or Mandated:

Voluntary: Firm initiated