Civco Medical Instruments Co. Inc.: Medical Device Recall in 2018 - (Recall #: Z-2946-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
CIVCO Needle Guide, REF 676-144, QTY 24, STERILE EO
Class II
The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide.
UDI 00841436107327, Lot numbers: A018067, A018067, A021026, A021026, A021226, A021226, A021226, A022447, A022447, A023876, A023876, A025137, A025137, A025137, A025137, A02560, A025606, A025606, A025606, A026609, A026609, A026609, A026609, A026610, A026610, A026610, A028159, A028159, A028645, A029293, A029293, A029293, A029383, A032036, A032036, A032036, A032036, A032036, A032036, A033139, A033139, A033139, A033139, A033153, A033153, A035009, A035009, A035009, A035010, A035010, A035010, A035010, A035010, A035568, A035568, A037150, A037231, A037231, A037231, A037585, A037585, A039417, A039417, A040376, A040376, A040839, A041444, A041444, A041444, A041723, A041723
United States, Austria, Belgium, Denmark, France, Germany, Poland, Spain, Switzerland and United Kingdom
Voluntary: Firm initiated