Clinical Diagnostic Systems: Medical Device Recall in 2018 - (Recall #: Z-1322-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
VITROS Immunodiagnostic Products TSH Reagent Pack
Class III
VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.
Lot 5470, expiry date: 24-May-2018 Lot 5430, expiry date: 01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017,
Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.
Voluntary: Firm initiated