Clinical Innovations, LLC: Medical Device Recall in 2020 - (Recall #: Z-2118-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICAL INNOVATIONS Manufacturer: Clinical Innovations, LLC* * *Reorder Number: UM750* * *LOT 181183* * *." Each device is accompanied by Instruction for Use which is similarly labeled.

Product Classification:

Class II

Date Initiated: May 30, 2019
Date Posted: June 3, 2020
Recall Number: Z-2118-2020
Event ID: 85144
Reason for Recall:

An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury.

Status: Terminated
Product Quantity: 280 devices
Code Information:

Model Number: UM750 Lot Number: 181183

Distribution Pattern:

International distribution to the country Japan.

Voluntary or Mandated:

Voluntary: Firm initiated