ConBio, a CynoSure Company: Medical Device Recall in 2013 - (Recall #: Z-1599-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Multilite Dye Handpieces, an optional accessory of the RevLite SI Laser. The Revlite SI laser is intended for General Dermatology, Dermatologic and General Surgical Procedures. Which includes treatments such as tattoo removal, pigmented lesion removal, removal or lightening of hair, skin resurfacing and reduction of red pigmentation in hypertropic and keloid scars The two Multilite Dye Handpieces (HPs) are optional handpieces that can be used with the Revlite SI Laser. Manufactured by: ConBio, A Cynosure Company 47733 Fremont blvd Fremont, CA 94538

Product Classification:

Class II

Date Initiated: May 9, 2013
Date Posted: July 3, 2013
Recall Number: Z-1599-2013
Event ID: 65407
Reason for Recall:

An inconsistency was identified related to electronic display identification of the optional Multilite Dye Handpieces of the RevLite SI Laser system that may confuse user.

Status: Terminated
Product Quantity: 12 (6 of model 659-1050; 6 of model 659-1100)
Code Information:

659-1050-585 nm Handpiece; Manufacturer numbers: 659-1050 / C13B1141 659-1050 / D13C1151 659-1050 / D13C1157 659-1050 / D13C1158 659-1050 / D13C1160 659-1050 / D13C1154 Model 659-1100 - 650 nm Handpiece; manufacturer numbers: 659-1100 / C13B1141 659-1100 / D13C1151 659-1100 / D13C1157 659-1100 / D13C1158 659-1100 / D13C1160 659-1100 / D13D1165.

Distribution Pattern:

Worldwide distribution: USA (nationwide) including states of: IL, ME, NM and VA; and countries of: Australia and Chile.

Voluntary or Mandated:

Voluntary: Firm initiated