ConvaTec, Inc: Medical Device Recall in 2019 - (Recall #: Z-1037-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM Product Usage: Urinary Catheters are sterile disposable devices intended to provide an intermittent pathway for drainage of fluids from the bladder. They are intended to be inserted intermittently through the urethra and up to the bladder enabling urine to drain. They are designed for short-term, intermittent transient use only and are not intended to be left in situ (there is no balloon on the tip to allow the catheter to be retained within the bladder). They are intended for single use only. They are intended to be removed once the bladder is fully drained. The goal of intermittent catheterization using Urology Catheters is to completely empty the bladder and prevent Urinary Tract Infections (UTIs).

Product Classification:

Class II

Date Initiated: January 9, 2019
Date Posted: March 27, 2019
Recall Number: Z-1037-2019
Event ID: 82039
Reason for Recall:

An internal assessment of packaging confirmed the potential for a breach in the sterile barrier.

Status: Ongoing
Product Quantity: 5,744,525 units total
Code Information:

Product Codes/REF Numbers: 505599, 505603; Lot Codes: 222348, 222349

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of IL, MI, PA, WI. and the countries of Algeria, Australia, Austria, Belgium, Canada, Croatia, Czechia, Czech Republic, Denmark, Egypt, Estonia, Finland, Faroe Islands, France, Germany, Greece, Hong Kong, Hungary, Iceland, Israel, Italy, Kenya, Korea, Kuwait, Lebanon, Myanmar, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Spain, Slovakia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, United States, Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated