Cook Inc.: Medical Device Recall in 2017 - (Recall #: Z-2540-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages
Class II
Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
Vue Optic Visualization Source, Catalog Number: FVO-150, GPN: G25343 Flexor Vue Deflecting Endoscopic System Catalog Number: FV-090075-150, GPN: G50972 Catalog Number: FV-090045-150, GPN: G34306
Domestic: FL IN LA MS NC TX Foreign:Argentina, Australia, Brazil, Canada, Czech Republic, Egypt, Estonia, France, Germany, Israel, Italy, Netherlands, Norway, Paraguay, Poland, Portugal, Russian Federation, Spain, Switzerland, Tunisia, Turkey, United Kingdom, VA/DOD:None
Voluntary: Firm initiated