Cook Inc.: Medical Device Recall in 2018 - (Recall #: Z-1461-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
Product Classification:
Class II
Date Initiated: March 22, 2018
Date Posted: April 25, 2018
Recall Number: Z-1461-2018
Event ID: 79713
Reason for Recall:
Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
Status: Terminated
Product Quantity: 2,267
Code Information:
All lots
Distribution Pattern:
Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Voluntary or Mandated:
Voluntary: Firm initiated