Cook Inc.: Medical Device Recall in 2018 - (Recall #: Z-1752-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Connecting Tube - Female to Male Luer Lock, HPCT8.8-60-M-FM-BNS Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
Class II
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
NS7881928, NS7898133
Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.
Voluntary: Firm initiated