Cook Inc.: Medical Device Recall in 2020 - (Recall #: Z-0973-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
C-Flex Double Pigtail Ureteral Stent Set, Global Product No. G14364 G14365 G14367
Class II
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
NS7953136 NS8169661 NS8180857 NS8169662 NS8180858 NS8180859 NS8180860
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Voluntary: Firm initiated