Cook Inc.: Medical Device Recall in 2020 - (Recall #: Z-0974-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151
Class II
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
8011027 8183322 8037660 8030233 8018054 7994816
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Voluntary: Firm initiated