Cook Incorporated: Medical Device Recall in 2022 - (Recall #: Z-1707-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RFSPC-35-180 ORDER NUMBER (GPN) G09608. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Class II
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
UDI-DI: (01)00827002096080- Lot Numbers: 14071906, 14091204, 14134663, 14134667, 14134669, 14134670, 14134673, 14134676, 14143872, 14143873, 14160997, 14184102, NS14143735, NS14152903
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Voluntary: Firm initiated