Cook Incorporated: Medical Device Recall in 2022 - (Recall #: Z-1715-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150 ORDER NUMBER (GPN): G56176. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Product Classification:

Class II

Date Initiated: July 19, 2022
Date Posted: September 21, 2022
Recall Number: Z-1715-2022
Event ID: 90710
Reason for Recall:

Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Status: Ongoing
Product Quantity: 865 units
Code Information:

UDI-DI: (01)00827002561762- Lot Numbers: 14112845, NS14062106, NS14080814, NS14089120, NS14118843, NS14139358, NS14142006, NS14144747, NS14148490, NS14149799, NS14153797, NS14172915

Distribution Pattern:

Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA

Voluntary or Mandated:

Voluntary: Firm initiated