Cook Incorporated: Medical Device Recall in 2022 - (Recall #: Z-1836-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
Newton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701 TSCFNA-35-145-3 G00561
Product Classification:
Class II
Date Initiated: August 16, 2022
Date Posted: October 5, 2022
Recall Number: Z-1836-2022
Event ID: 90793
Reason for Recall:
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Status: Ongoing
Product Quantity: 42 US
Code Information:
TSFNA-35-145 G00701 UDI-DI: 00827002007017 Lots/Expiration Date: 14798674 16-06-2027 TSCFNA-35-145-3 G00561 UDI-DI: 00827002005617 Lots/Expiration Date: 14835600 11-07-2027
Distribution Pattern:
Domestic distribution nationwide. Foreign distribution worldwide.
Voluntary or Mandated:
Voluntary: Firm initiated