Cook Incorporated: Medical Device Recall in 2022 - (Recall #: Z-1842-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Rosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623 THSCF-35-80-1.5-ROSEN G01774 THSCF-35-180-1.5-ROSEN G01264 THSCF-35-145-1.5-ROSEN G01261

Product Classification:

Class II

Date Initiated: August 16, 2022
Date Posted: October 5, 2022
Recall Number: Z-1842-2022
Event ID: 90793
Reason for Recall:

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Status: Ongoing
Product Quantity: 5041 US; 298 OUS
Code Information:

THSCF-35-220-1.5-ROSEN G01623 UDI-DI: 00827002016231 Lots/Expiration Date: 14818344 29-06-2027 14820965 30-06-2027 14849857 20-07-2027 THSCF-35-80-1.5-ROSEN G01774 UDI-DI: 00827002017740 Lots/Expiration Date: 14820960 30-06-2027 14832877 08-07-2027 14847945 19-07-2027 THSCF-35-180-1.5-ROSEN G01264 UDI-DI: 00827002012646 Lots/Expiration Date: 14820961 30-06-2027 14843133 15-07-2027 14847946 19-07-2027 THSCF-35-145-1.5-ROSEN G01261 UDI-DI: 00827002012615 Lots/Expiration Date: 14820962 30-06-2027 14820963 30-06-2027 14820964 30-06-2027 14828587 06-07-2027 14835364 11-07-2027 14835365 11-07-2027 14849859 20-07-2027

Distribution Pattern:

Domestic distribution nationwide. Foreign distribution worldwide.

Voluntary or Mandated:

Voluntary: Firm initiated