Cook Medical Llc: Medical Device Recall in 2024 - (Recall #: Z-0722-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Class II
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
Voluntary: Firm initiated