Cooper Surgical, Inc.: Medical Device Recall in 2012 - (Recall #: Z-0439-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy
Class II
Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle
All serial numbers
Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Czech Republic, Germany, Italy, Spain, France, United Kingdom, Japan, New Zealand, South Korea, Saudi Arabia, Sweden, Singapore, Slovenia, and Taiwan.
Voluntary: Firm initiated