Cordis Corporation: Medical Device Recall in 2021 - (Recall #: Z-2432-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598C
Product Classification:
Class I
Date Initiated: July 21, 2021
Date Posted: September 22, 2021
Recall Number: Z-2432-2021
Event ID: 88372
Reason for Recall:
Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement.
Status: Terminated
Product Quantity: 26025 units
Code Information:
all codes
Distribution Pattern:
Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe.
Voluntary or Mandated:
Voluntary: Firm initiated