Covidien LLC: Medical Device Recall in 2018 - (Recall #: Z-1130-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Situate Gauze RF and X-Ray detectable. Product Number: G0804-16P02CN-1 Situate premium cotton are single use cotton disposable that must be used with the Situate delivery system. The disposable have an attached RF Tag, which is an electrically passive inductor like device that is sewn into the green pouch that you will see on all Situate premium cotton. When activated by magnetic impulses from the scanning devices, the tag resonates a signal that allows for the detection of these cotton disposable.
Class II
This voluntary recall is being conducted due to the potential for failure to detect the affected Situate" gauze - RF and Xray detectable products by the Situate" detection console due to a manufacturing error.
Lot Number: 170704A UDI: 20884521700861
U.S.
Voluntary: Firm initiated