Covidien Llc: Medical Device Recall in 2020 - (Recall #: Z-1989-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Product Classification:
Class II
Date Initiated: April 30, 2020
Date Posted: May 20, 2020
Recall Number: Z-1989-2020
Event ID: 85629
Reason for Recall:
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Status: Terminated
Product Quantity: 8162 units
Code Information:
Lot Numbers: 1909308664, 1909320964 1933802964
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated