Covidien Llc: Medical Device Recall in 2020 - (Recall #: Z-1994-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P18XTSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Class II
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Lot Numbers: 1914133564
US Nationwide distribution.
Voluntary: Firm initiated