Covidien Llc: Medical Device Recall in 2020 - (Recall #: Z-2768-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Beacon EUS Access System Preloaded Access 90degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.

Product Classification:

Class II

Date Initiated: June 26, 2020
Date Posted: August 12, 2020
Recall Number: Z-2768-2020
Event ID: 86001
Reason for Recall:

This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.

Status: Ongoing
Product Quantity: 375 units
Code Information:

Model # DSA-090-01 UDI: 10884521722583 Lot #: F2510370X, F2510900X, F2512340X, F2512378X, F2512492X, F2512620X, F2513452X

Distribution Pattern:

world wide distribution

Voluntary or Mandated:

Voluntary: Firm initiated