Covidien Llc: Medical Device Recall in 2021 - (Recall #: Z-0353-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Puritan Bennett 980 Series Ventilator, 980X3JADIJJ
Product Classification:
Class I
Date Initiated: November 4, 2021
Date Posted: December 15, 2021
Recall Number: Z-0353-2022
Event ID: 88987
Reason for Recall:
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Status: Terminated
Product Quantity: 278 total
Code Information:
GTIN/UDI - 10884521172531 Serial Numbers: 35B1700393 35B1700419 35B1700420 35B1700427 35B1700428 35B1700429 35B1700540 35B1700564 35B1700568 35B1700573 35B1800002
Distribution Pattern:
Worldwide distribution - US Nationwide.
Voluntary or Mandated:
Voluntary: Firm initiated