Covidien Llc: Medical Device Recall in 2022 - (Recall #: Z-0557-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular. Product Numbers CH08-40-75GB and CH08-40-75US

Product Classification:

Class II

Date Initiated: December 17, 2021
Date Posted: February 9, 2022
Recall Number: Z-0557-2022
Event ID: 89418
Reason for Recall:

The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.

Status: Ongoing
Product Quantity: 1,994 devices
Code Information:

UDI: 10884521812253, 07290016745047, 10884521812291 For CFN CH08-40-75US: GTIN Batch 07290016745047 21904374 07290016745047 21904432 07290016745047 22000673 07290016745047 22000709 07290016745047 22000817 07290016745047 22000880 07290016745047 22001030 07290016745047 22001061 07290016745047 22001191 07290016745047 22001310 07290016745047 22001417 07290016745047 22001448 07290016745047 22001728 07290016745047 22002086 07290016745047 22002088 10884521812253 22002498 07290016745047 22002583 10884521812253 22002603 07290016745047 22003037 07290016745047 22100003 07290016745047 22100321 07290016745047 22100322 07290016745047 22101089 07290016745047 22101090 10884521812291 22102189

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.

Voluntary or Mandated:

Voluntary: Firm initiated