Covidien, LLC: Medical Device Recall in 2023 - (Recall #: Z-2008-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS
Product Classification:
Class II
Date Initiated: May 22, 2023
Date Posted: July 5, 2023
Recall Number: Z-2008-2023
Event ID: 92394
Reason for Recall:
Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.
Status: Ongoing
Product Quantity: 1 unit
Code Information:
UDI-DI: 10884521782211; Serial Number: 518503
Distribution Pattern:
International distribution in the country of Taiwan.
Voluntary or Mandated: