Covidien: Medical Device Recall in 2023 - (Recall #: Z-0169-2024)
See the recall detail below. You can also see other recalls from the same firm in 2023.
McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only
Class II
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
UDI: 10884521824386/ Lot # 22102001
US: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM NY OH OK OR PA SD TN TX UT VA WA WI WY OUS: Australia Austria Azerbaijan Bahamas Belgium Croatia Czech Republic Denmark Finland France French Guiana Germany Guadeloupe Hong Kong Hungary Ireland Israel Italy Kazakhstan Korea, Republic Of Malta Mayotte Netherlands New Zealand Norway Pakistan Poland Portugal Reunion Romania Spain Sweden Switzerland Turkmenistan United Arab Emirates
Voluntary: Firm initiated