Curbell Medical, Inc.: Medical Device Recall in 2014 - (Recall #: Z-2554-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.

Product Classification:

Class II

Date Initiated: August 5, 2014
Date Posted: September 3, 2014
Recall Number: Z-2554-2014
Event ID: 68961
Reason for Recall:

The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.

Status: Terminated
Product Quantity: Domestic: 112 units; Foreign: 211 units
Code Information:

Lot 0414

Distribution Pattern:

Nationwide, Canada, Ireland, Italy

Voluntary or Mandated:

Voluntary: Firm initiated