Cure Medical LLC: Medical Device Recall in 2014 - (Recall #: Z-0496-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Pediatric, 10Fr, Intermittent Urinary Catheters Model: P10 Pouch Label UPC (P10) *+M405P100C* Intermediate Box Label UPC (P10) *+M405P103F* Carton Label UPC (P10) *+M405P105H* The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. Intended to be used by pediatric males and females for the purpose of bladder drainage

Product Classification:

Class II

Date Initiated: November 11, 2014
Date Posted: December 10, 2014
Recall Number: Z-0496-2015
Event ID: 69737
Reason for Recall:

Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.

Status: Terminated
Product Quantity: 192 cases = 1,920 boxes = 57,600 each
Code Information:

Lot Number: 140417-3

Distribution Pattern:

Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated