Customed, Inc: Medical Device Recall in 2015 - (Recall #: Z-2645-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

ARTHROSCOPY PACK 3/Cs Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .

Product Classification:

Class II

Date Initiated: July 10, 2015
Date Posted: September 9, 2015
Recall Number: Z-2645-2015
Event ID: 71721
Reason for Recall:

Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Status: Terminated
Product Quantity: 75
Code Information:

Code:9002609 Lots: 141216307 exp. 12/31/15 150116554 exp. 1/31/16 150317853 exp. 5/31/16 150317852 exp. 5/31/16 150116796 exp. 5/30/16 150317854 exp. 5/31/16

Distribution Pattern:

Distributed Only in Puerto Rico.

Voluntary or Mandated:

Voluntary: Firm initiated